Development and Diagnostic Performance Evaluation of a Wireless Digital Fetal Heart Rate Monitoring System Using Ultrasonic Doppler with Digital Filter-Based Noise Reduction

Authors

  • Muhammad Ridha Makruf Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya
  • Roichatun Nashichah Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya
  • Lusiana Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya
  • Verry Angriawan Department of Medical Electronics Technology, Poltekkes Kemenkes Surabaya

DOI:

https://doi.org/10.33019/

Keywords:

Fetal Heart Rate Monitoring, Doppler Ultrasound, Wireless Communication, ESP32, Digital Signal Processing, Low-Pass Filter, Real-Time Monitoring

Abstract

Early fetal heart rate (FHR) monitoring is essential for assessing fetal well-being and enabling the early detection of fetal distress during pregnancy. However, existing fetal Doppler systems commonly rely on wired communication and remain susceptible to signal noise, limiting monitoring flexibility and measurement reliability. This study aimed to develop and evaluate a wireless fetal heart rate monitoring system with enhanced signal quality and reliable real-time communication. The proposed system integrates an ultrasonic Doppler sensor, ESP32 microcontrollers communicating through the ESP-NOW protocol, and digital signal processing using a Butterworth low-pass filter and a smoothing algorithm to suppress noise and improve signal clarity. System performance was evaluated using sampling frequencies ranging from 200 to 1000 Hz and baud rates from 300 to 250,000 bps. Communication reliability was assessed by transmitting 1,500 ADC packets for each sampling-frequency configuration and 1,000 ADC packets for each baud-rate configuration while monitoring packet sequence numbers to identify transmission loss. Experimental results showed that no packet loss was observed under all tested configurations. The developed system achieved BPM measurement errors ranging from 0.56% to 0.89% and improved signal quality, with the highest signal-to-noise ratio of 18.70 ± 0.17 dB obtained using a 0.5–20 Hz Butterworth filter. Higher sampling frequencies produced more detailed waveform representations, whereas higher baud rates improved transmission responsiveness without affecting waveform integrity within the tested range. These findings demonstrate that the proposed system provides an accurate, reliable, and portable solution for real-time wireless fetal heart rate monitoring.

Downloads

Download data is not yet available.

References

[1] Misnaniarti, N. Sariunita, and H. Idris, “Regional perinatal mortality differences in Indonesia: Evidence from Indonesian demographic health survey,” Public Heal. Pract., vol. 7, p. 100501, 2024, doi: https://doi.org/10.1016/j.puhip.2024.100501.

[3] M. B. Landon et al., Obstetrics: Normal and Problem Pregnancies E-Book: Normal and Problem Pregnancies. Saunders, 2020. [Online]. Available: https://books.google.co.id/books?id=92PRDwAAQBAJ

[4] F. G. Cunningham, K. J. Leveno, J. S. Dashe, B. L. Hoffman, C. Y. Spong, and B. M. Casey, Williams Obstetrics, Twenty-sixth Edition. McGraw Hill LLC, 2022. [Online]. Available: https://books.google.co.id/books?id=2HVKEAAAQBAJ

[5] S. A. Alnuaimi, S. Jimaa, and A. H. Khandoker, “Fetal Cardiac Doppler Signal Processing Techniques: Challenges and Future Research Directions.,” Front. Bioeng. Biotechnol., vol. 5, p. 82, 2017, doi: 10.3389/fbioe.2017.00082.

[6] E. H. Hon, “Fetal Dopplerometry Fetal heart rate Links Related articles,” pp. 1–2, 1958.

[7] M. B. Landon et al., Gabbe’s Obstetrics: Normal and Problem Pregnancies, 8th ed. Philadelphia, PA: Saunders, 2020.

[8] F. Ahmed, M. Ibrahimy, M. Mohd Ali, and E. Zahedi, “A portable recorder for long-term fetal heart rate monitoring,” Microprocess. Microsyst., vol. 26, pp. 325–330, Sep. 2002, doi: 10.1016/S0141-9331(02)00043-1.

[9] M. Garimella, A. Bandari, and I. Patel, “Microcontroller Based Fetal Heart Rate Monitoring using Intelligent Biosystem,” 2013. [Online]. Available: https://api.semanticscholar.org/CorpusID:12631341

[10] Y. Hirono et al., “The Approach to Sensing the True Fetal Heart Rate for CTG Monitoring: An Evaluation of Effectiveness of Deep Learning with Doppler Ultrasound Signals,” Bioengineering, vol. 11, no. 7, 2024, doi: 10.3390/bioengineering11070658.

[11] A. Nurfathoni, “RANCANG BANGUN FETAL DOPPLER NIRKABEL BERBASIS ESP32 DAN NRF24L01 UNTUK TRANSMISI AUDIO DAN NILAI FHR,” J. Ilm. Poli Rekayasa, vol. 21, p. 34, Oct. 2025, doi: 10.30630/jipr.21.1.427.

[12] A. A. Boatin et al., “Wireless fetal heart rate monitoring in inpatient full-term pregnant women: testing functionality and acceptability.,” PLoS One, vol. 10, no. 1, p. e0117043, 2015, doi: 10.1371/journal.pone.0117043.

[13] S. Murray, D. J. Fox, R. L. Coddington, and V. L. Scarf, “How does the use of continuous electronic fetal monitoring influence women’s experiences of labour? A systematic integrative review of the literature from high income countries,” Women and Birth, vol. 37, no. 4, p. 101619, 2024, doi: https://doi.org/10.1016/j.wombi.2024.101619.

[14] C. Eenkhoorn, T. Goos, J. Dankelman, A. Franx, and A. Eggink, “Evaluation and patient experience of wireless noninvasive fetal heart rate monitoring devices,” Acta Obstet. Gynecol. Scand., vol. 103, Jan. 2024, doi: 10.1111/aogs.14776.

[15] I. Campos, H. Gonçalves, J. Bernardes, and L. Castro, “Fetal Heart Rate Preprocessing Techniques: A Scoping Review,” Bioengineering, vol. 11, no. 4, 2024, doi: 10.3390/bioengineering11040368.

[16] R. Setyawati, P. C. Nugraha, H. G. Ariswati, and N. H. Ahniar, “An Improved Measurement Accuracy of Fetal Heart Rate using Digital Filter,” vol. 2, no. 3, pp. 136–142, 2020.

[17] H. Zhang, H. Zhao, and Z. Guo, “Artificial Intelligence-Based Atrial Fibrillation Recognition Method for Motion Artifact-Contaminated Electrocardiogram Signals Preprocessed by Adaptive Filtering Algorithm,” Sensors, vol. 24, no. 12, 2024, doi: 10.3390/s24123789.

[18] A. M. Maghfiroh, S. Y. Setiawan, M. Umer, and F. Mujahid, “Digital Filter Design to Reduce Motion Artifacts in Electrocardiogram Signals Based on IIR Filter,” vol. 16, no. 4, pp. 232–237, 2023.

[19] J. Wei, Z. Wang, and X. Xing, “A Wireless High-Sensitivity Fetal Heart Sound Monitoring System.,” Dec. 2020, Switzerland. doi: 10.3390/s21010193.

[20] S. Fatimah, “Optimization of heart rate measurement using fast Fourier transform modeling on electrocardiogram signals,” vol. 6, no. 2, pp. 113–120, 2026, doi: 10.59190/stc.v6i2.313.

Downloads

Published

18.10.2026

How to Cite

[1]
M. Ridha Makruf, R. Nashichah, Lusiana, and V. Angriawan, “Development and Diagnostic Performance Evaluation of a Wireless Digital Fetal Heart Rate Monitoring System Using Ultrasonic Doppler with Digital Filter-Based Noise Reduction”, JurnalEcotipe, vol. 13, no. 2, pp. 202–211, Oct. 2026, doi: 10.33019/.

Most read articles by the same author(s)